Page 1 - Operating instructions EN

IKA® EUROSTAR 20 high speed controlEUROSTAR_10201420000003967Operating instructions EN

Page 2

10Menu structureStirringSpeed Limit ………………………………………………………………………… 6000 rpm Torque Limit ……………………………………………………………..………… 20 Ncm Torque Display ………………

Page 3 - Note for Canada (IC)

11Menu (details) StirringSpeed Limit:The ”Speed Limit” menu allows the user to set the desired maximum upper speed limit for the EUROSTAR 20 high spee

Page 4 - Safety instructions

12DisplayIn the “Display“ menu the user can specify what information will be displayed on the main screen.Note: If the ”Torque”option is activated, by

Page 5 - Unpacking

13Sample for editing the program:PROGRAM 1SegNo.rpm Timehh:mmInterm.ModeEdit Delete Insert12345 0 00:00 NoeditPROGRAM 1SegNo.rpm Timehh:mmInterm.Mode

Page 6

14SafetyTime Out:In the “Time Out“ menu, the user can set a time limit. This time limit applies if there is a communication failure between the EU-ROS

Page 7 - Securing

15micro B micro AUSB micro BUSB micro AConnection capability: Wireless Controller to the EUROSTAR station:Connection capability: EUROSTAR stat

Page 8 - Switching on the instrument

16orA9-pin RS 232 9-pin RS 232USB micro B USB Amicro B USB AMaintenance and cleaningThe instrument is maintenance-free. It is only subject to

Page 9 - 00:00:00

17WarrantyIn accordance with IKA® warranty conditions, the warranty period is 24 months. For claims under the warranty please contact your local deale

Page 10

18EUROSTAR station + Wireless Controller (WiCo)Speed range under nominal load rpm 0 / 150 – 6000Speed adjustment SteplessSpeed display TFT / Wireless

Page 11 - ∞ / CW

IKA® - Werke GmbH & Co.KGJanke & Kunkel-Str. 10D-79219 StaufenTel. +49 7633 831-0Fax +49 7633 [email protected]

Page 12 - √) disappears

2Fig. 31234Fig. 2Fig. 8Fig. 7Fig. 5 Fig. 6HIJGGTUQSUXIGZHGWVQBMONCLAJIHGFEPDEUROSTAR stationWireless Controller(WiCo)PRS 232USBTemp. SensorFig. 4NOPRF

Page 13 - PROGRAM 1 Detail

3ContentsEC-Declaration of conformity 3Note for USA (FCC) 3Note for Canada (IC) 3Explication of warning symbols 3Safety instructions 4Correct u

Page 14 - Interfaces and outputs

4• The instrument can only be disconnected from the mains supply by pulling out the mains plug or the connector plug.• The socket for the mains cord m

Page 15

5• The machine must only be opened by trained specialists, even dur-ing repair. The instrument must be unplugged from the power sup-ply before opening

Page 16 - USB micro B USB A

6Accessories must be assembled according to the following assembly instructions (Fig. 2 to Fig. 7).If above conditions are met, the instrument is read

Page 17 - Warranty

7SecuringSecuring the extension arm to the overhead stirrerDiagram (see Fig. 2)Ensure that the extension arm is fitted securely.Vibration may cause

Page 18 - Technical data

8Useful information The EUROSTAR 20 high speed control stirrer is controlled via a Wireless Controller (WiCo). If the Wireless Controller is attached

Page 19

9Working screen at the time of delivery:Explanation of symbols on the working screen:The symbols displayed change depending on the status and settings