Ika labworldsoft User Manual Page 136

  • Download
  • Add to my manuals
  • Print
  • Page
    / 158
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 135
- 136 -
0HWURKPS+0HWHU
pH Meter with automatic temperature compensation.
Determination of a primary measured value together with current temperature.
0HD VXUHGYDOXHV
1. pH value
2. Temperature [°C]
3. Potential [mV]
1RWHV  
Please see Metrohm Instructions for Use for a detailed description of the instrument.
For proper communication, the pH Meter must be connected using a suitable RS232 cable
Metrohm order No.: 6.2125.010). The RS232 settings at the pH Meter must be set to the following
parameters:
EDXGUDWHGDWDELWVWRSELWSDULW\QRQHKDQGVKDNH+:V56FRQWURO21
These are default values of the pH Meter configuration under: &21),*56VHWWLQJV
The response time of this instrument is 100 ms for each activated measured value. If both
channels are activated, the minimum sampling time is increased to 200 ms. If a lower sampling
time was chosen by the user, this value will be automatically adjusted by labworldsoft with the
beginning of the measurement.
Page view 135
1 2 ... 131 132 133 134 135 136 137 138 139 140 141 ... 157 158

Comments to this Manuals

No comments