Ika labworldsoft User Manual

Browse online or download User Manual for Software Ika labworldsoft. IKA labworldsoft User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 158
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews

Summary of Contents

Page 1 - PDQXDO9

- 1 - ODEZRUOGVRIWPDQXDO9±,.$:HUNH*PE+&R.*http://www.ika.de

Page 2 - 7DEOHRI&RQWHQWV

- 10 - The icon bar contains a number of action buttons (Windows standard symbols) for direct selection of the following functions:  Start or stop

Page 3

- 100 - ,.$:HUNH'DWDFRQWURO,2 The IO 2 (,nput - 2utput) module from the IKA Laboratory Technology company provides for 8 digital inputs (

Page 4

- 101 - ,.$:HUNH'DWDFRQWURO'$ With '$ from the IKA Laboratory Technology company laboratory devices with analog inputs can be

Page 5

- 102 - 'HYLFHFRQILJXUDWL RQDQDORJG H YLFHV ODEZRUOGVRIW device configuration makes it possible to integrate laboratory devices with anal

Page 6

- 103 - 1. Call up the parameters window by double clicking on the icon. The parameter window then appears. In the basic setting analog port 1 is a

Page 7 - ([LWLQJSURJUDP

- 104 - ,.$:HUNH'DWDFRQWURO'& With the Datacontrol DC 2 laboratory instruments or instruments of other manufacturers with analog

Page 8 - 0LQLPL]HDFWLRQE XWW RQ

- 105 - appears. 2. Select the desired analog instrument from the list by clicking on the list entry. The selected list entry appears inverted. 3.

Page 9 - ,FRQEDU

- 106 - ,.$:HUNH'70The DTM 11 made by IKA Labortechnik is an all-purpose PT-100 resistance thermometer with a measuring range from -100 t

Page 10 - 7RROEDU

- 107 - ,.$:HUNH'=0FRQWURO The DZM control is intended for a wide range of speed measuring tasks. The instrument operates either according

Page 11

- 108 - ,.$:HUNH(85267$532:(5FRQWUROYLVF The EUROSTAR POWER control-visc is a high-torque stirrer for use up to the high-viscosity range. The

Page 12 - 6HOHFWLQJRSWLRQ

- 109 - ,.$:HUNHIH[,.$ The solids extractor is a device used to extract all types of solid matter quickly and fully automatically. It is based on

Page 13

- 11 - /DERUDWRU\LQVWU XP H QWSymbol for a laboratory instrument. &RQWUROXQLWSymbol of the "Rated Value" control unit. &RQWURO

Page 14 - $FWLRQEXWWRQV

- 110 - Cycles 1. n=Number of operating cycles: Here, you can enter how often the process described above is to be repeated in the experiment seque

Page 15 - *HQHUDWLQJVLJQDOIORZFKDUW

- 111 - The illustration explains the fexIKA processing sequence: Legend: TH, TK, tS, tF: See the above text

Page 16 - 6DYLQJVLJQDOIORZFKDUW

- 112 - ,.$:HUNH.+6 The KHS makes it possible to operate the fexIKA (solid-fluid serial extractor) not only as a unit with fixed cycles, but al

Page 17 - 6LJQDOIORZFKDUW

- 113 - ,.$:HUNH6KD NH U .0 Processor controlled horizontal shaker. &RQWURO1. Input path for set speed. 0HDVXUHGYDOXHV1. Set spe

Page 18 - Positioning functional units

- 114 - ,.$:HUNH.6.6.6FRQWURO Family of processor-controlled horizontal shakers with different maximum loading and speed ranges of

Page 19 - Removing connections

- 115 - ,.$:HUNH/DERUDWRU \3LORW The laboratory pilot primarily consists of a single-stage high-performance device for the continual dispersal o

Page 20 - - 20

- 116 - 7KHUPRV WDW/7&RQWURO Highly precise, processor-controlled laboratory thermostat for direct and indirect thermostating. &RQWURO

Page 21 - 6HWWLQJSDUDPH WHUV

- 117 - ,.$:HUNH/7 Highly precise, processor-controlled laboratory thermostats for direct and indirect tempering. &RQWURO1. Input path f

Page 22 - Result windows

- 118 - ,.$:HUNH3$GLJLWDO Peristaltic pumps are used as dosing, feed or suction pumps in all laboratory fields. A large selection of interchange

Page 23

- 119 - ,.$:HUNH5(7FRQWUROYLVF The RET control-visc is a magnetic stirrer with heating function and viscosity trend recognition. The instrumen

Page 24 - Setting sampling time

- 12 - 6HWWLQJSDUDPH WHUVZLWKGLDORJER[The following illustration shows typical dialog boxes of labworldsoft Proceed as follows: x Click on

Page 25 - ,PS RUW DQW

- 120 - ,.$:HUNH5(7FRQWUROYLVFVDIHW\FRQWURO The 5(7FRQWUROYLVF VDIHW\FRQWUROmagnetic stirrerby IKA Werke is a magnetic stirrer with heat

Page 26 - Opening file

- 121 - 3. Time-out actions The5(7FRQWUROYLVF VDIHW\FRQWUROmagnetic stirrerby IKA Werke offers internal monitoring by labworldsoft® or the c

Page 27 - File Manager

- 122 - ,.$:HUNH9,6&2./O&.9. With the VISCOKLICK VK 250/600 stirrers can be expanded to torque measuring instruments. The viscos

Page 28 - File info

- 123 - ,.$:HUNH9;5FRQWURO Processor-controlled horizontal shaker with speed range up to 2000 1/min. &RQWURO1. Input path for set speed.

Page 29 - 8VHU0DQDJHPHQW

- 124 - ,VPDW H F 3XPSV  Peristaltic pumps are used as dosing, feed or suction pumps in all laboratory fields. A large selection of interchangeabl

Page 30

- 125 - -XODER/DERUDWRU\7KHUPRVWDWV Highly precise, processor-controlled laboratory thermostats for direct and indirect tempering. &RQWURO

Page 31 - &RQWUROXQ L WV

- 126 - .HUQ%DODQFHV Precision laboratory balances from the Kern company. &RQWURO1. The balance does not have any control inputs. 0HDVXU

Page 32 - 5DWHGYDOXHFRQWUROXQLW

- 127 - .1))ORGRV67(3'265& A Stepdos pump is a microprocessor-controlled membrane dosing pump for liquids. It is used for the q

Page 33 - /DWFKFRQWUROXQLW

- 128 - ..QLFN&RQGXFW RPHWHU  The laboratory conductometer 703 from the Knick company is used for electrolytic conductivity measurement i

Page 34

- 129 - .QLFNS+±0HWHU The laboratory pH Meter 765 is used to measure pH and Redox. 0HDVXUHGYDOXHV1. pH value in pH. 2. Electrode voltag

Page 35

- 13 - x Use the "Tab" key to reach the desired option group. Press the "space bar". &RQWUROILHOGVCertain options can be

Page 36 - 7ULJJHUFRQWUROXQLW

- 130 - .QLFN3RUWDPHVV&RQG The Conductometer Portamess 913 Cond is used to measure condcutivity. The device may be optionally operated w

Page 37 - 5HODLV FRQWURO X QLW

- 131 - .QLFN3RUWDPHVV2[\ The Oxymeter Portamess 913 Oxy is used to measure oxygen concentration. 0HDVXUHGYDOXHV1. Temperature in °C. 2

Page 38

- 132 - .QLFN3RUWDPHVVS+ The pH Meter Portamess 913 pH is used to measure pH 0HDVXUHYDOXHV1. pH value 2. Electrode potential [V] 3. T

Page 39 - 6SHFLI\ LQ JVFDOH

- 133 - /DXGD/DERUDWRU\7KHUPRVWDWV Highly precise, processor-controlled laboratory thermostats for direct and indirect tempering. &RQWURO1

Page 40 - 'HOHWLQJVXSSRUWSRLQWV

- 134 - 0HWURKP'RVLPDW Dosing device for pulse-free dispensing and exact pipetting. 2SHUDWLRQLQWKUHHGLIIHUHQWPRGHV '26: Dos

Page 41 - 0RYLQJVXSSRU WSRLQWV

- 135 - 1RWHV    Please see Metrohm Instructions for Use for a detailed description of the instrument. For proper communication, the Dosimat

Page 42 - 5HDGILOHVFRQWUROXQLW

- 136 - 0HWURKPS+0HWHU pH Meter with automatic temperature compensation. Determination of a primary measured value together with current t

Page 43 - 7LPHU&RQWURO%ORFN

- 137 - 0HWURKPS+,RQ0HWHU pH/Ion Meter with automatic temperature compensation. Determination of a primary measured value together with c

Page 44

- 138 - 0HWWOHU7ROHGR%DODQFH V Precision laboratory balances from the Mettler-Toledo company. &RQWURO1. The balance does not have any con

Page 45

- 139 - 0HWWOHU7ROHGR&RQGXFWRPHWHU0& The Conductometer MC 126 is used to measure condcutivity. 0HDVXUHGYDOXHV1. Conductivity 2.

Page 46 - 3:0&RQWU RO%ORFN

- 14 - $FWLRQEXWWRQVAction buttons directly initiate an action. In labworldsoft a number of different action buttons are provided from which the f

Page 47

- 140 - 0HWWOHU7ROHGR&RQGXFWRPHWHU0& The Conductometer MC 226 is used to measure condcutivity. 0HDVXUHGYDOXHV1. Conductivity 2.

Page 48

- 141 - 0HWWOHU7ROHGR2[\JHQ0HWHU02 The Oxygen-Meter MO 128 is used to measure oxygen concentration. 0HDVXUHGYDOXHVOxygen concentration [

Page 49

- 142 - 0HWWOHU7ROHGRS+0HWHU030$ The laboratory pH-Meter MP 123 is used to measure pH The laboratory pH-Meter MA 130 is used to me

Page 50 - 0RQRIORS&RQWURO%ORFN

- 143 - 0HWWOHU7ROHGRS+0HWHU030303 The laboratory pH-Meter MP 225, MP 227, MP 230 are used to measure pH. 0HDVXUHGYDOXHV1.

Page 51

- 144 - 0/79DFXXPFRQWU ROOHU39. Vacuum controller for accurate and ecological control of vacuum pumps and water jet units. &RQWURO1.

Page 52 - &RXQWHU)XQFWLRQ%ORF N

- 145 - 1HVODE7KHUPRVWDW57( Highly precise, processor-controlled laboratory thermostat for direct and indirect thermostating. &a

Page 53 - To set the averaging method

- 146 - 3RO\6FLHQFH 7KHUPRVWDW'7& Highly precise, processor-controlled laboratory thermostats for direct and indirect tempering. &R

Page 54 - ,QFUHDVLQJDYHUDJLQJ

- 147 - 6DUWRULXV %DODQFHV Precision laboratory balances from the Sartorius company. &RQWURO1. The balance does not have any control inputs

Page 55 - - 55

- 148 - 6FDOWHF%DODQFHV Precision laboratory balances from the Scaltec company. &RQWURO2. The balance does not have any control inputs. 0

Page 56 - Multi-channel operation

- 149 - 6LJPD&HQWULIX JHV Microprocessor controlled centrifuge available in an non-cooled and cooled version. &RQWURO1. Input path for

Page 57

- 15 - %DVLFRSHUDWLQJVHTXHQFH*HQHUDWLQJVLJQDOIORZFKDUWThe main components of the signal flow chart are generally the laboratory instruments,

Page 58

- 150 - 7HODE3XPSVMicroprocessor-controlled diaphragm metering pump for liquids and gases. The corresponding input and output paths are marked by

Page 59

- 151 - 9DFXXEUDQG9DFXXPFRQWUROOHU&9& The vacuumcontroller CVC 2000 is used to control precise vacuum of membrane pumps. &RQWUR

Page 60 - )XQFWLRQEORFNYLVFRVLW\

- 152 - 9DFXXEUDQGYDFXXPJDXJH'95 The DVR 5 is a fully electronic, versatile vacuum gauge for a wide range of measurements from 1100 to

Page 61

- 153 - <HOORZOLQH6KDNHU Family of processor-controlled horizontal shakers with different maximum loading and speed ranges of 30 - 800 1/min.

Page 62 - %HDULQJIULFWLRQFXUYH

- 154 - $XWRPDWLFDGMXVWPHQWRIWKHVFDQQLQJUDWHUnder normal conditions the minimum scanning rate per activated device measured value amounts to

Page 63 - 77//HYHO

- 155 - ,QGH[A Activating input and output paths for connections ... 20 Admi

Page 64 - 6HOHFWLQJUHVXOWXQLWV

- 156 - G General information starting and exiting program...

Page 66 - 'LJLWDOG LVSOD\

- 158 - T Table of Content ...

Page 67 - 6HWWLQJIRQW

- 16 - &RQQHFWLQJIXQFWLRQDOXQLWVIn the third step the signal flow is defined by connecting the input and output paths. This completes the sig

Page 68 - *UDSKGLVS OD\

- 17 - 6LJQDOIORZFKDUWThe basis for the control, measuring and evaluation operations on the screen is the signal flow chart, in which the physica

Page 69 - 6FDOLQJ<D[LV

- 18 - Designing the tool bar customer-specifically The tool bar can be assigned as required with all of the function blocks available in labworldV

Page 70 - Assign second curve

- 19 - 1RWH  The number fields are used as anchor points for the connections (arrows). Each output path (right-hand number field) of a function

Page 71 - Specify colors

- 2 - 7DEOHRI&RQWHQWV*(77,1*72.12:7+(6<67(0 

Page 72 - 2IIOL QH JUDSKGLVSOD\

- 20 - 2. Answer the question with <HV. The dialog box closes and the connection arrow is removed from the working area.

Page 73 - 'HILQLQJ <VFDO H

- 21 - 6HWWLQJSDUDPH WHUVLabelling functional units Specific descriptions with a maximum of 13 characters can be entered in the corresponding text

Page 74 - <;JUDSKLFGLVSOD\

- 22 - Setting and labelling channel number The "Averaging" and "Arithmetic" functional units and part of the result units can o

Page 75 - $KOERUQ$O PHPR

- 23 - Opening control or result window Proceed as follows: x Double click on the symbol of the window to be opened. x Reducing control or resu

Page 76

- 24 - &RQWUROOL Q JPHDVXU H PHQWVHTXHQFHStarting and stopping measurement The measuring sequence can either be controlled manually with the

Page 77

- 25 - For example, in the basic setting with a measurement duration of 30 min. and a sampling time of 500 ms (two values per second) a total of 3,6

Page 78

- 26 - 5. etc. 1RWHAfter opening the dialog box, Channel 1 is already set for entering the description. &RQWUROOL Q JPHDVXU H PHQWVHTXHQFH

Page 79

- 27 - If a new signal flow chart is to be generated, previously saved configurations can also be used. By saving the old configuration under a new

Page 80

- 28 - In order to be able to make optimal use of this function it is advisable to fill in the corresponding Information Fields in the measurement s

Page 81

- 29 - 8VHU0DQDJHPHQWODEZRUOGVRIW makes it possible to specify various hierarchical groups for access to the measurement sequence schedules via u

Page 82 - )OXLG)/06 

- 3 - 7,0(5&21752/%/2&. 3,&a

Page 83 - )ULWVFK$QDO\VHWWH3UR

- 30 - For this reason the Administrator should point out to all of the other users that no other passwords that are used anywhere else should be us

Page 84

- 31 - &RQWUROXQ L WVThe following chapter contains a detailed description of the control units with all parameter settings. In addition to un

Page 85 - *)/)UHH]HUIUHH]LQJFDELQHW

- 32 - 5DWHGYDOXHFRQWUROXQLWThe Rated value control unit represents a slide controller with which the controlling variables, such as temperat

Page 86 - *)/6KDNHU

- 33 - /DWFKFRQWUROXQLW Latch The latch can be used to freeze values in place. If a High level (5) is present on the control input of the latc

Page 87 - *)/6KDNLQJ,QFXEDWRU

- 34 - 6ZLWFKEXWWRQFRQW U ROXQLW Functions can be reset or switched over with the 6ZLWFKHU. This includes, for example, the switching over of

Page 88 - +DDNH7KHUPRVWDW

- 35 - 3XVKEXWWRQ  F RQWUROXQLW Functions can be reset or switched over with the Push button. This includes, for example, the switching over o

Page 89

- 36 - 7ULJJHUFRQWUROXQLWThe trigger type and the limits of the range to be monitored may be specified using the trigger range. With the rang

Page 90

- 37 - 5HODLV FRQWURO X QLWThe module switches an input to an output as a function of a control input or interrupts the signal flow. The modul

Page 91 - ,QIXVLRQ0RGHO0RGHO

- 38 - 5DP SIXQFW LRQFRQWUROXQLWWith the Ramp function variables such as temperature, speed etc. can be automatically controlled. For example

Page 92 - +HLGROSK3XPSV

- 39 - 5DP SHGLWRUThe illustration shows the ramp editor window with the graph field and the current scale. The ramp function represents a tempera

Page 93

- 4 - *)/6+$.(5 

Page 94 - +HUPOH&HQWULIXJHV

- 40 - (QWHULQJVXSSRUWSRLQW VJUDSKLFDOO\  Proceed as follows: 1. Specify the 7LPHSRVLWLRQ of the new support points (2): Position the mouse p

Page 95 - +XEHU/DERUDWRU\7KHUPRVWDWV

- 41 - 0RYLQJVXSSRU WSRLQWV Proceed as follows: 1. Position the mouse pointer in the box of the support point to be moved. 2. Move the box to

Page 96

- 42 - 5HDGILOHVFRQWUROXQLWWith the Read files control unit complex measurement sequences can be displayed repeatedly in a simple manner or t

Page 97 - ,QIRUV0LQLWURQ

- 43 - 7LPHU&RQWURO%ORFN The 7LPHU control block generates up to 8 control signals with TTL-Pegel (0V / 5V). The amount of time for the cha

Page 98

- 44 - 3,'&RQWUROOHU &RQWURO%ORFN The 3,'&RQWUROOHU control block is a regulator circuit with an input for the 6HW9DOX

Page 99 - ,QIRUV6KDNH U 

- 45 - A 0D[LPXPDQGRU0LQLPXP2XWSXW9DOXH can be specified via the 0DQLSXODWHG9DULDEOH/LPLW. These values are not exceeded or undershot, even i

Page 100

- 46 - 3:0&RQWU RO%ORFN The 3:0 control block is a pulse spacing modulator with which data values can be digitized IO-Modul ) and displayed

Page 101

- 47 - 0XOW LFKDQQHO2SHUDWLRQThe ports should identifiable for multichannel operation. The ports may be identified for this purpose. Carry out t

Page 102 - &RQILJXULQJ'$

- 48 - %RROHDQ) X QF WLRQV&RQWURO%ORFN With the Boolean control block links, Boolean mathematical calculations can be carried out with in

Page 103 - Assigning Analog Devices

- 49 -  0XOW LFKDQQHO2SHUDWLRQThe ports should identifiable for multichannel operation. The ports may be identified for this purpose. Carry o

Page 104 - Assigning analog instruments

- 5 - ,.$:(5.(9,6&2.//&.9. ,.$:(5.(9;5

Page 105

- 50 - 0RQRIORS&RQWURO%ORFN The 0RQRIORS control block is a WLPHUZKLFKLVWULJJHUDEOHVWDUWDEOHYLDDFRQWUROLQSXWIf a change of the lo

Page 106 - ,.$:HUNH'70

- 51 - 1XPHULFDO%ORFN 'HULYDWLYH,QWHJUDO The 'HULYDWLYH,QWHJUDO control block is used to integrate data or for calculation of i

Page 107 - ,.$:HUNH'=0FRQWURO

- 52 - &RXQWHU)XQFWLRQ%ORF N The &RXQWHU function block can perform counting functions for input signals in the TTL-Bereich such asSRVLW

Page 108

- 53 - )XQFWLRQDOXQLW IRUDYHUDJLQJWith the Averaging functional unit, resulting result values xi can be averaged with two different methods: T

Page 109 - ,.$:HUNHIH[,.$

- 54 - %ORFNE\EORFNDYHUDJLQJMean values from n consecutive result values xi are generated, for example, for n = 5 of i = 1 to 5, i = 6 to 10, i

Page 110

- 55 -

Page 111

- 56 - )XQFWLRQDOXQLW IRUDULWKPHWLF With the Arithmetic functional unit three function groups are available:  Arithmetic operations with RQH

Page 112 - ,.$:HUNH.+6

- 57 - 4. Enter the &KDQQHOQDPH in the text field, i.e. make an entry for each channel selected. 5. Confirm the entries with 2..

Page 113 - ,.$:HUNH6KD NH U .0

- 58 - $ULWKPHWLFRSHUDWLRQVZLWKRQHRSHUDQGDQGFRQVWDQW&KDQQHOE\FKDQQHORSHUDWLRQVZLWKDFRQVWDQWFDUHSRVVLEOHZLWKWKLVIXQFWLRQJURXS

Page 114

- 59 -  Proceed as follows: 1. Open the parameter window by double clicking on the symbol. 2. If necessary, enter the 0RGXOHQDPH in the text f

Page 115 - ,.$:HUNH/DERUDWRU \3LORW

- 6 - <(//2:/,1(6+$.(5  $8720$7

Page 116 - 7KHUPRV WDW/7&RQWURO

- 60 - )XQFWLRQEORFNYLVFRVLW\This functional group allows calculation of the viscosity. The viscosity is derived from - the speed [rpm] Input 1,

Page 117 - ,.$:HUNH/7

- 61 - 'HILQLWLRQRIVWLUULQJHOHPHQWA mathematical model (performance characteristic) of the stirring element described by the parameters a0,

Page 118 - ,.$:HUNH3$GLJLWDO

- 62 - %HDULQJIULFWLRQFXUYHSpeed-dependent bearing friction (e.g. laboratory reactor LR 2000 ) should be eliminated in order to ensure precise t

Page 119 - ,.$:HUNH5(7FRQWUROYLVF

- 63 - 77//HYHOTTL (7ransistor-7ransistor-/ogic) levels define the voltage values for distinguishing a logic level 0 from a logic level 1. The vol

Page 120

- 64 - 5HVXOWXQLWVThe following chapter contains a detailed description of the result units with all their parameter settings. In addition to the

Page 121

- 65 - :ULWLQJILO H VWith the Write files result unit results from one or more channels can be continuously written to a specified directory on t

Page 122

- 66 - 'LJLWDOG LVSOD\With the digital display the results of a channel can be displayed with variable resolution (decimal places). For the

Page 123 - ,.$:HUNH9;5FRQWURO

- 67 - 6HWWLQJEDFNJURXQGFRORUAfter selecting the Background action button in the parameter window of the digital display, the editor for setting

Page 124 - ,VPDW H F 3XPSV 

- 68 - *UDSKGLVS OD\With the graph display the results of a maximum of eight channels can be displayed over time simultaneously. The number of ch

Page 125

- 69 - 'HILQLQJ WLPHDQG<VFDOHVThe time axis is dependent on the current measurement duration and must be set separately (menu bar of re

Page 126 - .HUQ%DODQFHV

- 7 - General information, starting and exiting program The labworldsoft program is operated with the mouse or the keyboard in the same manner as a

Page 127

- 70 - 1RWHIn the basic setting the Y-axis is assigned the unit rpm (speed) with a value range from 0 to 2,000. In addition, the channel is provid

Page 128

- 71 - Measured values can be read out at any point on the curve with vertical cursor lines. The left-hand cursor line forms the intersection (y1/t1

Page 129 - .QLFNS+±0HWHU

- 72 - 2IIOL QH JUDSKGLVSOD\With the "Offline" graph display result unit controlling variables can be recorded "offline",

Page 130

- 73 - 'HILQLQJ <VFDO HThe Y-axis can be assigned a maximum of four scales in accordance with the selected number of channels. To improve

Page 131 - .QLFN3RUWDPHVV2[\

- 74 - <;JUDSKLFGLVSOD\Measuring data for completed measurement cycles can be displayed using the Y/X graphic display result block. This all

Page 132 - .QLFN3RUWDPHVVS+

- 75 - /DERUDWRU\GHYLFHV$KOERUQ$O PHPR The $KOERUQ$OPHPR is a all-purpose measuring instrument which, depending on the type, accommodates up to

Page 133

- 76 - &KULVW5RWDU\9DFXXP&RQFHQWUDWRU Microprocessor controlled centrifuge with cooling and vacuum operation. &RQWURO1. Input pat

Page 134 - 0HWURKP'RVLPDW

- 77 - &RUQLQJS+  LRQPHW H U  pH meter for two independent measuring points with temperature compensation. 0HDVXUHGYDOXHV1. pH valu

Page 135

- 78 - (KUHW/DERUDWRU\FRQWUROOHUJUDGR[35 The laboratory automatic controllers grado 9x1 / PR2 are used for device control in various labo

Page 136 - 0HWURKPS+0HWHU

- 79 - the following values: JUDGR[%DXGVWDUWELWGDWDELWVVWRSELWSDULW\QRQHKDQGVKDNHQRQH An RS485-RS232 converter is al

Page 137 - 0HWURKPS+,RQ0HWHU

- 8 - ODEZRUOGVRIWPDL QDSSOLFDWLRQZLQGRZAfter starting the program the main application window always appears. It contains both standardized Win

Page 138 - 0HWWOHU7ROHGR%DODQFH V

- 80 - (\HOD%DWK&LUFXODWRURU)UHH]HU1&%17%3&&03) Highly precise, processor-controlled Bath, Circulator or Freezer for tem

Page 139

- 81 - (<(/$2YHQ,QFXEDWRURU)XUQDFH:)21'2926/7, Highly precise, processor-controlled Oven, ncubator or Furnace &RQWURO 1.

Page 140

- 82 - )OXLG)/06  The )/06 is a stirring motor with a powerful torque for use up to the "high viscosity" range. The maximum torque

Page 141

- 83 - )ULWVFK$QDO\VHWWH3URVertically oscillating laboratory sieving device for precise separation and classification of grain fractions. &am

Page 142

- 84 - *HUKDUGW6KDNHU/DERVKDNH/652/DERVKDNH/652 Gerhardt Shaker &RQWURO1. Input path for set speed. 0HDVXUHPHQWYDOXHV1. S

Page 143

- 85 - *)/)UHH]HUIUHH]LQJFDELQHW GFL sub-zero freezing cabinets with from 30 to 500 liters of effective space, and GFL sub-zero freezers with f

Page 144

- 86 - *)/6KDNHU GFL-Shaker &RQWURO2. Input path for set speed. 0HDVXUHPHQWYDOXHV3. Set speed in 1/min. 4. Actual speed in 1/min. 127

Page 145

- 87 - *)/6KDNLQJ,QFXEDWRU Circular shaker in incubation cabinet. &RQWURO1. Input path for set speed. 2. Input path for set temperature.

Page 146

- 88 - +DDNH7KHUPRVWDW Highly precise, processor-controlled laboratory thermostat for direct and indirect thermostating. &RQWURO1. Input pa

Page 147 - 6DUWRULXV %DODQFHV

- 89 - +DDNH7KHUPRVWDWV'&'/VHULHV Highly precise, processor-controlled laboratory thermostat for direct and indirect thermostati

Page 148 - 6FDOWHF%DODQFHV

- 9 - For reducing the window to an icon. 0D[LPL]HDFWLRQEXWW RQFor increasing the window to its maximum size. 0HQXEDUDQGSXOOGRZQPH Q XVThe

Page 149 - 6LJPD&HQWULIX JHV

- 90 - +DDNH7KHUPRVWDW3KRHQL[/LQH Highly precise, processor-controlled laboratory thermostat for direct and indirect thermostating. &RQWURO

Page 150 - 7HODE3XPSV

- 91 - +DUYDUG6\U LQJH3XPS3 +',QIXVLRQ0RGHO0RGHO High precision syringe pump for low to medium backflow pressures. &RQWURO

Page 151

- 92 - +HLGROSK3XPSV Peristaltic pumps are used as dosing, feed or suction pumps in all laboratory fields. A large selection of intercha

Page 152

- 93 - +HLGROSK/DERUDWRU\6WLUUHU5=5FRQWURO (OHFWURQLFODERUDWRU\VWLUUHUV&RQVWDQWO\KLJKWRUTXHRYHUHQWLUHVSHHGUDQJH

Page 153 - <HOORZOLQH6KDNHU

- 94 - +HUPOH&HQWULIXJHV Microprocessor controlled centrifuges from the Hermle company: &RQWURO1. Input path for set temperature. 2. I

Page 154

- 95 - +XEHU/DERUDWRU\7KHUPRVWDWV Highly precise, processor-controlled laboratory thermostats for direct and indirect tempering. &RQWURO1.

Page 155

- 96 - ,OPYDF9DFXXPFRQWUROOHU9&= Vaccumcontroller is used to control pressure of membrane pumps. &RQWURO1. Input path for set vacu

Page 156

- 97 - ,QIRUV0LQLWURQ Shaker with incubation hood &RQWURO1. Input path for set speed [1/min] 2. Input path for set temperature [°C] 

Page 157

- 98 - ,QIRUV0XOWLWURQ8SSHU0LGGOH/RZHU8QLW Shakers with up to three independently controllable shaker levels with tempering. &R

Page 158

- 99 - ,QIRUV6KDNH U  Shakers with up to three independently controllable shaker levels and tempering via an incubation hood. &RQWURO1. Inpu

Comments to this Manuals

No comments