7
Fig. 6
• Filling and draining:
- Before lling the uid into the bath, open the venting
grid as indicated in following gure.
- Check and make sure that the drain valve is closed (the
stop position in clockwise direction, see Fig. 7).
Note:
Please note information in section “Commis-
sioning”.
- Connect the mains plug and turn on the device with
mains switch (2
, see Fig. 1
).
- The low level warning message appear on the display.
Meanwhile, the bottom LED segment of the lights up
in red color to indicate the low fluid level
(see Fig. 8)
.
- Open the lling lid (5
, see Fig. 1
) and ll uid to the
bath.
Stop position
Drain valve
Drain port
Fig. 7
Note: Pay attention to the fluid
level information on the display and
fluid level indicator (see Fig. 8).
Bracket
- To drain the uid from the bath, connect a hose to the
drain port and turn the drain valve in counter clockwise
direction with a straight screwdriver.
Fig. 9
Red (Low Level)
Red (High Level)
Low Level
High Level
Fig. 8
NOTICE